Peak Voltage Voltmeter | MTS-PV10; digital multi-function peak voltmeter applies standard 19 inch and it has the functions such as AC/DC automatic distinguish, signal measurement, breakdown peak value protection and HV cross-line fast protection. The precision is 0.2/0.5 class.
MTS-PV10; applies DSP core computing and digital signal processing technology. It can identify the breakdown point precisely through real-time signal waveform analysis and amplitude and frequency calculations, maintain the max. voltage value when the break down happens and protect immediately.
Anti-EMC design is used in the instrument, which enables the instrument can work reliably in high electric field and magnetic field.
TECHNICAL PARAMETERS
Altitude | ≤ 1000m |
Ambient Temperature | -5℃~+40℃ |
Daily Average Temperature | ≤20℃ |
Max. Relative Humidity | ≤90℃ (25℃) |
Power Supply Waveform is sine wave, Waveform Deviation | <3% |
With a Reliable Earthling Spot, Earthing Resistance | <0.5Ω |
Signal Measuring Precision | 0.2/0.5 (Uin>1V) |
Signal Measuring Range | AC Signal: 40Hz-400Hz Input Voltage :0-700V Effective Value or DC: 0-1000V |
Input Impedance | >1MΩ |
Over Voltage Relay Protection Interface Capacity | Current : <1A Voltage : <250V |
Relay Protection Response Time | <20ms |
Power Supply | 220V/20VA |
Surrounding Temperature | -15 ℃-55 ℃, no condensation |
Signal Interface | Standard BNC signal terminal., standard R16 signal terminal |
AC Dividing Ratio K1 Setting Range | 0.01-99999.99 |
AC Dividing Ratio K2 Setting Range | 0.01-99999.99 |
Over Voltage Setting Range | 0.01-99999.99kV |
Break Down Limit Setting Range | 0.01-99999.99kV |
GENERAL INFORMATION
What is Peak Voltage?
Peak voltage is the peak or highest value of voltage for any voltage waveform.
Importance of Peak Voltage
The effects of peak voltage can damage the AC Motor and cause premature wear in the motor insulation.