Multifunctional CE Tester is suitable for routine testing of IEC 60065, IEC 60335, IEC 60601, IEC 60745, IEC 60884, IEC 61010, IEC 61029, IEC 61558 , IEC 60598-1etc.
It is designed to perform high voltage test, insulation resistance test, earth continuity test, leakage current test, low start and optionally power input test (with MT-6BATE model) in a single device.
Multifunctional CE Tester Technical Parameters:
EARTH CONTINUITY TEST
1.1 Current: 1A-30A AC Accuracy: 5% reading
1.2 Earth continuity: 10 ~ 200 mΩ (10A-25A)
10 ~ 600 mΩ (1A-10A)
Accuracy: 5% reading
1.3 Test time: 1 – 255 s
INSULATION RESISTANCE TEST
2.1 Test voltage: 100V-1000V DC, Accuracy: 5% reading
2.2 Insulation resistance: 1MΩ ~ 500MΩ, Doğruluk: Accuracy: 10% reading
2.3 Test time: 1-255 s
2.4 Device Protection
When the output current exceeds the normal range of the device, it will be evaluated by the program as a test failure caused by overcurrent and an “overcurrent failure” warning will appear on the screen.
HIGH VOLTAGE TEST
3.1 Test voltage: 100-4000VAC, Accuracy: 5% reading
3.2 Leakage current: 0.10mA ~ 12.00mA, Accuracy: 5% reading
3.3 Test time: 1-255 s
3.4 Device Protection
When the output current exceeds the normal range of the device, it will be evaluated by the program as a test failure caused by overcurrent and an “overcurrent failure” warning will appear on the screen.
When the test current is >12mA, software protection is activated and the test is finished and an “overtravel” warning is displayed.
LEAKAGE CURRENT TEST
4.1 Test voltage: 10-300V AC, Accuracy: + 3% reading and 2% range
4.2 Test current: 10-20000µA, Accuracy: + 3% reading and 2% range
4.3 Test time: 1-255 s
4.4 Device Protection
When the test current is> 22000 µA, the software protection is activated and the test is finished and an, “overtravel” warning is displayed.
4.5 Insulation mode: Regulator isolation, test voltage adjustment
POWER INPUT TEST ( For Model MT-6BATE)
Function | Measuring Range | Accuracy |
Voltage | 10.0~300.0V | (0.3% reading + 0.2% range) |
Current | 0.010-20.00A | (0.3% reading + 0.2% range) |
Power Input | 1.0~6000W | Cos φ > 0.5, (0.3% + 0.2% range) cos φ = 0.5, 0.5% + (+0.5% range) |
Power Factor | cos φ: 0.2-1 | ±0.02 |
6. PA TEST (POWER PARAMETER MEASUREMENT)
Alarm | Power upper/lower limit alarm; | |
Settings of power upper/lower limit | Range: 0.00W~6000W. Resolution: 0.01W /0.1W /1W, Error: ± (0.1%× setting +0.1%× range) | |
Active power measurement | Range: 0.10W~6.000kW. Resolution: 0.01W /0.01W /0.1W /0.001kW Accuracy: >PF 0.5: ± (0.1%× reading +0.1%× range) PF≤0.5: ± (0.4%× reading +0.1%× range) | |
Voltage measurement | Range: 5.00V~300.0V, peak factor: ≤1.6. Resolution: 0.01V /0.1V; Accuracy: ± (0.1%×reading +0.1%× range), 45Hz≤f≤65Hz | |
Current measurement | Range: 0.00mA~25.00A , peak factor: ≤1.6. Resolution: 0.01mA /0.1mA /0.001A /0.01A Accuracy: ± (0.1%× reading +0.1%× range), 45Hz≤f≤65Hz | |
Power factor measurement | Range: ± (0.100~1.000). Resolution: 0.001 Accuracy: ±0.01 (voltage/current amplitudes are greater than 10% of the corresponding ranges) | |
Frequency measurement | Range: 45.00Hz ~65.00Hz. Resolution: 0.01Hz. Accuracy: ± (0.1%× reading ) | |
Duration | Range: 0, (0.5~999.9) s, 0: infinite. Resolution: 0.1s. Accuracy:± 0. 1%× setting + 2 word) |