Main features
* Comply with CCC, IEC/EN,VDE, BS, UL, JIS standard requirements.
* Latest international standard: 500VA AC Hi-pot capacity(AC 5KV/100mA), short-circuit output current exceeds 200mA. Built-in optional leakage current test network MD card, meeting requirements of different industrial standards. Maximum test current of Ground Bond resistance is up to 64A,
* 6 in 1: AC Hi-pot, Insulation Resistance, Ground Bond resistance, Leakage Current, Running Power, Low-Voltage Start,
* Standby power consumption is optional for AN9651F(F)and AN9651S(F), minimum is low to 0.50W, accuracy is 0.2%,
* Simple operation: LED backlit screen display menu operation, Chinese/English languages, socket test fixture,
* Flexible configuration: Built in isolated transformer, AC frequency conversion power supply, frequency conversion power supply and industrial PC combination,
* Satisfy latest requirements: Follow the production line developing trends of comprehensive, fast, automation and information. Satisfy quality management, 5s management and refit management requirements.
Technical Parameters:
AC Hipot test | ||
Booster rated output capacity | 500VA(5000V/100mA) | |
Booster maximum short-circuit current | >200mA | |
Output waveform | Sine wave, <3%(pure resistance,5000V/1mA),<3%(From empty load to full load) | |
Output voltage setting | Range | 500~5000V |
Resolution | 5V | |
Accuracy | ±(3%×setting+10V) | |
Frequency setting | Range | 50Hz/60Hz |
Accuracy | ±0.1%×setting | |
Alarming limits setting | Current upper-limit | 0.10~99.99mA |
Current lower-limit | 0.00~10.00mA | |
Compensation current setting | Range | 0.00~10.00mA, automatic testing, compensation can be set to ON/OFF |
Time setting | Range | 1~999s |
Ramp up time range | 1~100s | |
Accuracy | ±(0.1%×setting+1count) | |
Breakdown current alarm | Range | |
Accuracy | ||
Insulation Resistance Test | ||
Rated output capacity | 2VA(1000V/2mA) | |
Output voltage regulation | <3%(From no load to full load) | |
Output voltage ripple | <1% | |
Output voltage setting | Range | DC 500Vor1000V |
Accuracy | ±(3%×setting+3counts) | |
Alarming limits setting | Resistance upper-limit range | 0.0~99.9 MΩ,100~2000MΩ(If upper-limit is set to 0, it will not make adjudgment for upper limit) |
Resistance lower-limit range | 0.3~99.9 MΩ, 100~2000MΩ | |
Time setting | Range | 1~999s |
Accuracy | ±(0.1%×setting+1count) | |
Insulation resistance measurement | Range | 0.0~99.9 MΩ, 100~2000 MΩ |
Accuracy | ≤200MΩ: ±(3%×reading+3counts) | |
>200MΩ:±(8%×reading+8counts) | ||
Ground Bond Resistance test | ||
No load output voltage | <12V | |
Output current setting | Range | 3~30A |
Accuracy | ±(3%×setting+0.2A) | |
Current range instruction | Optional: 3 ~ 64A test current | |
Current frequency setting | Range | Sine waveform 50Hz/60Hz |
Accuracy | ±0.1%×setting | |
Test time setting | Range | 1~999s |
Accuracy | ±(0.1%×setting+1count) | |
Result display mode | Ground resistance or voltage | |
Ground resistance display mode | ||
Alarming limits setting | Resistance upper-limit range | 3A≤output current≤10A:10~600 mΩ |
Resistance lower-limit range | 0~100 mΩ | |
Compensation resistance setting | Range | 0~200 mΩ, automatic testing, compensation can be set to ON/OFF |
Ground bond resistance measurement | Range | 10~600 mΩ |
Accuracy | ±(3%×reading+ 5counts) | |
Ground voltage display mode | ||
Alarming limits setting | Voltage upper-limit range | 30~7500 mV |
Voltage lower-limit range | 0~3000 mV | |
Compensation voltage setting | Range | 0~1000 mV, automatic testing, compensation can be set to ON/OFF |
Ground bond resistance measurement | Range | 10~7500 mV |
Accuracy | ±(3%×reading+ 50mV) | |
Leakage Current test | ||
Leakage current test type setting | Single phase load, static/dynamic leakage current | |
MD network selection | Standard IEC 60335-1 figure 4 network;other test network optional | |
Alarming limits Setting | Current upper-limit range | 0.050~9.999mA |
Current lower-limit range | 0.000~5.000mA | |
Leakage current compensation setting | Range | 0.000~1.000mA, automatic testing, compensation can be set to ON/OFF |
Time test setting | Range | 1~999s |
Accuracy | ±(0.1%×setting+1count) | |
Leakage current measurement | Range | 0.050~9.999mA |
accuracy | DC~10kHz:±(3%×reading+0.010mA) | |
Output voltage measurement | Range | 60~280V(phase voltage) |
accuracy | ±(0.5%×reading+2counts) | |
Run Power Only forAN9651S(F)/ AN9651F(F)/ AN9651B(F) | ||
Alarming limits setting | Power upper-limit range | 6~6000W |
Power lower-limit range | 0~6000W | |
Test time setting | Range | 1~999s |
Accuracy | ±(0.1%×setting+1count) | |
Voltage measurement | Range | 60~280V(phase voltage) |
Accuracy | ±(0.5%×reading+2counts) | |
Current measurement | Range | 0.030~3.999A, 4.00~25.00A |
Accuracy | ±(0.5%×reading+2counts) | |
Active power measurement | Range | 30.0~199.9W:±(0.5%×reading+5W) |
Accuracy | 200~6000W(or2000W):±(0.5%×reading+30W) |