Electrical Safety Tester Desktop

Desktop Electrical Safety Tester, model AN9640A (F) and AN9640B (F) are designed for AC Hipot, Insulation Resistance, Ground Resistance, Leakage Current, Operating Power and Low Voltage Start-up tests.

DEVICE INTRODUCTION VIDEO

Order Now
Description

Electrical Safety Tester Desktop are designed for AC Hi-pot, Insulation Resistance, Ground Bond Resistance, Leakage Current, Running Power and Low-Voltage Start up tests.

These testers are used to perform the following electrical safety tests required in international standards:

– Ground Bond Resistance Test,

– Insulation Resistance Test,

– AC Hi-pot Test,

– Leakage Current Test.

Related standards: IEC60335, IEC60950, IEC60601, IEC61010, IEC60598, IEC60204, etc.

Features

  • ground bond test up to 64A AC with resistance measurement up to 600mΩ
  • insulation test with DC Voltage from 1V up to 1000 VDC (1V steps),500VA AC Hi-pot capacity(AC 5KV/100mA),short-circuit output current exceeds 200mA.Built-in optional leakage current test network MD card, meeting requirements of different industrial standards Maximum test current of Ground Bond resistance is up to 64A
  • leakage current measurement in single or three phases, according to all current international standards
  • AC Hi-pot, Insulation Resistance, Ground Bond resistance, Leakage Current, Running Power, Low-Voltage Start.
  • power measurement and cos φ measurement
  • Satisfy latest requirements: Follow the production line developing trends of comprehensive, fast, automation and information. Satisfy quality management,5s management and refit management requirements.
  • GPIB, RS232, Ethernet, 0-10V, PLC available
  • ESRS software for quality purposes, traceability and professional reporting

Electrical Safety Tester Desktop  Technical Parameters:

AC Hipot test

Booster rated output capacityAN9640A(F)E500VA(5000V/100mA)
AN9640B(F)E100VA(5000V/20mA)
Booster maximum short-circuit current>200mA
Output waveformSine wave, <3%(pure resistance,5000V/1mA),<3%(From empty load to full load)
distortion and regulation
Output voltage settingRange500~5000V
Resolution5V
Accuracy±(3%×setting+10V)
Frequency settingRange50Hz/60Hz
Accuracy±0.1%×setting
Alarming limits settingCurrent upper-limit0.10~99.99mA(20.00mA for AN9640B/C(F))
Current lower-limit0.00~10.00mA
Compensation current settingRange0.00~10.00mA, automatic testing, compensation can be set to ON/OFF
Time settingRange1~999s
Ramp up time range1~100s
Accuracy±(0.1%×setting+1count)
Breakdown current alarmRange0.10~99.99mA(20.00mA for AN9640B/C(F))
Accuracy±(3%×reading+3counts)

Insulation Resistance test

Rated output capacity2VA(1000V/2mA)
Output voltage regulation<3%(From no load to full load)
Output voltage ripple<1%
Output voltage settingRangeDC 500Vor1000V
Accuracy±(3%×setting+3counts)
Alarming limits settingResistance upper-limit range0.0~99.9 MΩ,100~2000MΩ(If upper-limit is set to 0, it will not make adjustment for upper limit )
Resistance lower-limit range0.3~99.9 MΩ, 100~2000MΩ
Time settingRange1~999s
Accuracy±(0.1%×setting+1count)
Insulation resistance measurementRange0.0~99.9 MΩ, 100~2000 MΩ
Accuracy≤200MΩ: ±(3%×reading+3counts)
RemarksExcept 500V and 1000V voltage points, other voltage accuracy are all 10%. Insulation resistance accuracy is under temperature 040  humidity 40/2075% RHotherwise the accuracy will change.

Ground Bond Resistance test

No load output voltage<12V
Output current settingRange3~30A
Accuracy±(3%×setting+0.2A)
Current frequency settingRangeSine waveform 50Hz/60Hz
Accuracy±0.1%×setting
Test time settingRange1~999s
Accuracy±(0.1%×setting+1count)
Result display modeGround resistance or voltage
Ground resistance display mode:
Alarming limits settingResistance upper-limit range3A≤output current≤10A:10~600 mΩ
Resistance lower-limit range0~100 mΩ
Compensation resistance settingRange0~200 mΩ, automatic testing, compensation can be set to ON/OFF
Ground bond resistance measurementRange10~600 mΩ
Accuracy±(3%×reading+ 5counts)
Ground voltage display mode:
Alarming limits settingVoltage upper-limit range30~7500 mV
Voltage lower-limit range0~3000 mV
Compensation voltage settingRange0~1000 mV, automatic testing, compensation can be set to ON/OFF
Ground bond resistance measurementRange10~7500 mV
Accuracy±(3%×reading+ 50mV)

DUT power control

DUT input power requirementsLeakage current, running power and voltage start test, user shall provide the correct input power input to safety tester, recommend to use AN97 frequency conversion power supply
Input voltage, frequency0~300V(phase voltage), 45~65Hz
Isolation transformer output voltage ratioLeakage current test1.06 times rated voltage
Running Power test1.00 times rated voltage
Voltage start test0.85 times rated voltage
DUT CapacitySingle phase, AN9640A(F) load capacity is 6 KVA
AN9640B(F)/AN9640C(F) load capacity is 2KVA
Over current protectionMax current 30A, after the over current 5s,automatically cut off the isolation transformer.

Interface

Remote interfaceStandard, foot pedal(start/stop)
Alarming lamp interfaceStandard, 3 colour light(test pass/fail)
Communication interfaceStandard RS232,internet, RS485, GPIB optional
97 Power supply control interfaceStandard,combine with AN97 series frequency conversion power
PLC interfaceOptional, start、stop、test、pass、fail output
Printer interfaceOptional
Gross WeightAbout 50kg
DimensionWith packing600(W)x385(H)x820(D)
Without packing426(W)x 178(H)x 600(D)