Electrical Safety Comprehensive Tester AN9640BV3(F)
- The AC Hipot Test allows you to stress your product’s insulation with high voltage to ensure it will limit current flow should a fault occur.
- Output 100 mA of AC Hipot/Dielectric Withstand test current. Ainuo AN Series meets the 200 mA short circuit requirement for 500 VA Hipot Testers.
- The DC Hipot Test allows you to stress your product’s insulation and limit leakage current output by running a DC dielectric withstand test.
- The Ground Bond,Verify the integrity of your product’s protective earth ground conductor to prevent electric shock should a fault occur.
- The The Ground Continuity,Verify the presence of your product’s protective earth ground conductor to prevent electric shock should a fault occur.
- The Insulation Resistance,Determine the total resistance of your product’s insulation.
- The Leakage Current,Ensure your product’s leakage current remains at acceptable levels during 8 different fault conditions.
- The Functional Run,Verify that your product is working properly before shipment.
DESCRIPTION OF PRODUCT
The new Ainuo AN96 series safety tester has integrated seven functions (seven-in-one: AC withstand voltage, DC withstand voltage,insulation resistance, ground bond resistance, leakage/contact current, operating power, low voltage start-up, providing safety test solutions for household appliances, energy-saving lamps, medical devices, information equipment, electric vehicles, photovoltaic energy, power electronics and other industries, and innovative solutions for various intelligent automated tests.
PRODUCT FEATURES
- Precise: basic Accuracy: 1%; Accuracy of electrical parameters: 0.2%;
- Fast: Accurate testing within 0.5s, 0.1s fast switching, high-speed communication, high efficiency.
- Integration: integrate safety performance and electrical performance seven functions in one, one-key start for a station;
- Intelligent: Built-in 100 groups of test conditions, barcode recognition, automatic matching of test program.
- Automation: RS232(standard)/RS485/GPIB/LAN ports, barcode scanning interface, frequency conversion power control interface, easy to form an automated test system.
AN9640BV(F) Electrical Safety Comprehensive Analyzer | ||
Features | Technical indicator | |
AC WITHSTAND VOLTAGE TEST | ||
Rated output capacity | 500VA(5000V/100mA),short-circuit current more than 200mA | |
Output voltage | range/accuracy | 100~5000V,±(2%×setting +5 counts) |
Output frequency | range/accuracy | 50Hz or 60Hz,±0.1%×setting |
Alarm current setting | Upper limit range/accuracy | 0.10~40.00mA,±(2%×setting +5counts) |
Lower limit range/accuracy | 0.10~40.00mA,±(2%×setting +5counts) | |
Time setting | Range of test time | 0.5~999.9s(0 is infinite) |
Range of ramp-up time | 0.1~999.9s | |
Range of ramp-down time | 0.1~999.9s | |
Time setting accuracy | ±(1%×setting + 1count) | |
Measurement meter | Voltage range/accuracy | 0.01~5.00kV,±(1%× reading +1 count) |
Frequency range/accuracy | 50Hz or 60Hz,±0.1%×reading | |
Current range/accuracy | 0.10~40.00mA,±(1%×setting +5 counts)( | |
Time range/accuracy | 0.1~999.9s,±(1%×reading +1 count) | |
Output voltage waveform/degree of distortion/ Regulation rate | Sine wave,<2%(pure resistance load),
<(1%×setting+5V)(from no-load to full-load) |
|
Initial voltage setting | (0%~50%)×Output voltage setting | |
Current compensation setting | 0.000~ 10.00mA,Outomatic | |
Arc detection | 1~9 levels, level-9 corresponds to the most sensitive, 0 is Off | |
DC WITHSTAND VOLTAGE TEST | ||
Output voltage | Range/accuracy | 100~6000Vdc,±(1%×setting+5 counts) |
Alarm current setting | Upper limit range/accuracy | 0~10000μA,±(1%×setting +5 counts) |
Lower limit range/accuracy | 0~999.9μA,±(1%×setting +5 counts) | |
Time setting | Range of test time | 0.5~999.9s(0 is infinite) |
Range of ramp-up time | 0.4~999.9s | |
Range of ramp-up time | 1.0~999.9s(0 is off) | |
Accuracy | ±(1%×setting +1 count) | |
Measurement meter | Voltage range/accuracy | 0.01~6.00kVdc,±(1%×reading +5 count) |
Current range/accuracy | 0.0~10000µA,±(1%×reading +5 counts) | |
Time range/accuracy | 0.1~999.9s,±(1%×reading +1 count) | |
Rated output capacity | 60VA(6000Vdc/10mA) | |
Output adjustment level | ± (1%×setting +5 V), empty load to full load | |
Compensating current setting | 0 ~200.0μA,automatic, manual | |
Initial voltage setting | (0%~50%)× setting Output voltage | |
Lower limit current for charging | (0~350)μA,automatic, manual | |
Discharging time | ≤200ms | |
Arc detection | 1~9 levels, level-9 corresponds to the most sensitive, 0 is Off | |
INSULATION RESISTANCE TEST | ||
Output voltage | Range/accuracy | 100~2500Vdc,±(1%×setting +5V) |
Alarm resistance setting | Range of upper/lower limit | 1~50000 MΩ
The upper limit includes no upper setting |
Accuracy | Range:
1 MΩ~50.00GΩ Resolution: 0.001MΩ /0.01MΩ /0.1MΩ /0.001GΩ /0.01GΩ Error: 100V~499V:0.100MΩ~2.000GΩ, ± (5%× reading +2 words) 500V~2500V: 0.100MΩ~999.9MΩ, ± (2%× reading +2 words) 1.000GΩ~9.999GΩ: ± (5%× reading +2 words) 10.00GΩ~50.00GΩ: ± (15%× reading +2 words) |
|
Time setting | Range of ramp-up time | 0.1~999.9s |
Range of delay time | 0,0.5~999.9s(0 is infinite) | |
Range of test time | 1.0~999.9s,off | |
Range of ramp-down time | 1.0~999.9s,off | |
Setting accuracy | ±(1%×setting +1 count) | |
Measurement meter | Voltage range/accuracy | 100~2500Vdc,±(2%×reading +5 counts) |
Resistance range/accuracy | Range:
1 MΩ~50.00GΩ Resolution: 0.001MΩ /0.01MΩ /0.1MΩ /0.001GΩ /0.01GΩ Error: 100V~499V:0.100MΩ~2.000GΩ, ± (5%× reading +2 words) 500V~2500V: 0.100MΩ~999.9MΩ, ± (2%× reading +2 words) 1.000GΩ~9.999GΩ: ± (5%× reading +2 words) 10.00GΩ~50.00GΩ: ± (15%× reading +2 words) |
|
Time range/accuracy | 0.1~999.9s,±(1%×reading +1 count) | |
Lower limit current for charging | 0~3.500μA,automatic, manual | |
GROUNDING RESISTANCE TEST | ||
Rated output | Max output is 32A, max testing resistance 600mΩ,open-circuit voltage <12V | |
Output current setting | Range/accuracy | 2.0~32.0A±(1%×setting +2 counts) |
Output voltage setting | Range/accuracy | 3.0~10.0V ,±(1%×setting +2 counts) |
Output frequency setting | Range/accuracy | sine,50Hz or 60Hz,±0.1%×setting |
Alarm limit setting | Range of resistance upper/lower limit | Range: (0.1~600) mΩ, (2.0~10.6) A; (0.1~R) mΩ, (10.7~32.0) A. Resolution: 0.1/1 mΩ
Where R= (6400 / current setting ) mΩ |
Error | Error: ± (1%× setting +2 words) | |
Compensation resistance setting | Range | 0~200 mΩ,automatic measurement, compensation on/off |
Test time setting | Range/accuracy | 0.5~999.9s(0=infinite);±(1%×setting +1 count) |
Measurement meter | Current range/accuracy | 2.0~32.0A,±(1%×reading+2 counts) |
Voltage range/accuracy | 3.0~7.5V ,±(1%×reading +2 counts) | |
Resistance range/accuracy | Range: (10.0~99.9) mΩ. (100~600) mΩ. Resolution: 0.1/1 mΩ.
Error: ± (1%× reading +2 words) |
|
Time range/accuracy | Range: 0, (0.5~999.9) s, 0: infinite. Resolution: 0.1s. Accuracy: ± 0. 1%× setting + 2 word) | |
LEAKAGE CURRENT TEST | ||
Test Mode | Single phase load, leakage current at operating temperature (dynamic) and leakage current at non operating temperature (static), L G, N G, AUTO (G L, G N); Require external isolation transformer or inverter power to provide the required test voltage and frequency. | |
Working power state | Polarity: On, Off, Auto; Zero line: On, Off; Earth line: On, Off | |
Built in human network | Polarity switch: | |
Setting of alarm voltage | Range: 20.0V~300.0V AC. Resolution: 0.1V. Error: ± (0.4%× setting +0.1%× ra
nge) |
|
Voltage measurement | Range: 0.0V~300.0V, 45Hz~65Hz
Accuracy: 20.0V~300.0V: ± (0.4%× reading +0.1%× range) |
|
Load current | Upper limit: 30A | |
Settings of Touch/Leakage
current upper/lower limit (rms) |
Range: 0.0μA~12.00mA. Resolution: 0.1μA /1μA /0.01mA
Error: DC, 15Hz≤f≤100kHz: ± ( 1.5 %× setting 10 words) 100kHz f≤1000kHz: ±5%× setting |
|
Touch/leakagecurrent
measurement (rms) |
0.0μA~999.9μA: D C, 15Hz≤f≤100kHz: ± ( 1.5 %× reading 10 words)
100kHz f≤1000kHz, 10.0μA 999.9μA: ±5%× reading 1000μA~7999μA: DC, 15Hz≤f≤100kHz: ± ( 1.5 %× reading 10 words) 100kHz f≤1000kHz, 10μA 7999μA: ±5%× reading 8.00mA~12.00mA: DC: ± ( 1.5 %× reading 10 words) 15Hz≤f 1000kHz, 0.01mA~12.00mA: ±5%× reading |
|
LC Compensation | Range: 0.000~1.000mA, Auto measurement, On or Off. | |
Duration | Range: 0, (1 ~999.9), 0: infinite. Resolution: 0.1s. Accuracy:± 0. 1%× setting + 2 word), (for testmode AUTO (G L, G N): half of the duration) | |
DC input impedance | 2kΩ ± 1% (GB12113Figure 4) | |
Input impedance | ≤100kHz 5%; >100kHz 10% | |
Frequency response | Accuracy: same as current measurement error | |
PA TEST (POWER PARAMETER MEASUREMENT) | ||
Alarm | Power upper/lower limit alarm; | |
Settings of power upper/lower
limit |
Range: 0.00W~6000W. Resolution: 0.01W /0.1W /1W,
Error: ± (0.1%× setting +0.1%× range) |
|
Active power measurement | Range: 0.10W~6.000kW. Resolution: 0.01W /0.01W /0.1W /0.001kW
Accuracy: >PF 0.5: ± (0.1%× reading +0.1%× range) PF≤0.5: ± (0.4%× reading +0.1%× range) |
|
Voltage measurement | Range: 5.00V~300.0V, peak factor: ≤1.6. Resolution: 0.01V /0.1V;
Accuracy: ± (0.1%×reading +0.1%× range), 45Hz≤f≤65Hz |
|
Current measurement | Range: 0.00mA~25.00A , peak factor: ≤1.6. Resolution: 0.01mA /0.1mA /0.001A /0.01A
Accuracy: ± (0.1%× reading +0.1%× range), 45Hz≤f≤65Hz |
|
Power factor measurement | Range: ± (0.100~1.000). Resolution: 0.001
Accuracy: ±0.01 (voltage/current amplitudes are greater than 10% of the corresponding ranges) |
|
Frequency measurement | Range: 45.00Hz ~65.00Hz. Resolution: 0.01Hz. Accuracy: ± (0.1%× reading ) | |
Duration | Range: 0, (0.5~999.9) s, 0: infinite. Resolution: 0.1s. Accuracy:± 0. 1%× setting + 2 word) | |
ST TEST (LOW VOLTAGE START TEST) | ||
Settings of current upper/lower
limit |
Range: (0.00~25 .00) A. Resolution: 0.01A. Error: ± (0.1%× setting +0.1%× | |
Voltage measurement | Range: 5.00V~300.0V,peak factor: ≤1.6. Resolution: 0.01V /0.1V;
Accuracy: ± (0.1%× reading +0.1%× range), 45Hz≤f≤65Hz |
|
Current measurement | Range: (0.02~25 .00) A, peak factor: ≤1.6. Resolution: 0.01A
Accuracy: ± (0.1%× reading +0.1%× range), 45Hz≤f≤65Hz |
|
Duration | Range: 0,(0.5~999.9) s, 0: infinite. Resolution: 0.1s. ± 0. 1%× setting + 2 word) | |
FUNCTIONAL INTERFACE | ||
Remote controller interface | Standard, optional foot switch (start up) | |
Alarm lamp connector | Standard, optional three-color alarm lamp ( test, pass, alarm) | |
Communication port | Standard, network port, RS232,RS485, GPIB
one of four options, default RS232 |
|
Barcode interface | Standard, RS232 interface, barcode automatic recognition and programming match | |
U disk storage connector | Standard, support standard U disk connector, test data automatic storage | |
Switching value interface | Standard, can output the switching value of startup, stop, test, pass and alarm | |
Printer connector | Standard, support needle printer an d micro-printer | |
Rear panel test output interface | Standard, rear panel output interface of high-voltage and grounding tests | |
GENERAL SPECIFICATIONS | ||
Verification and calibration environment | 23℃±5℃,(45~75)%RH,little dust | |
Operating environment | 0 ºC ~40ºC,,(20~90)%RH,little dust | |
Storage environment | -10℃~50℃,(5~95)%RH,24h | |
Input Power | AC 220V±10%, 50Hz±5%, 10A | |
Power consumption of system | Max 550W | |
Weight | About 45kg | |
Dimension (mm, without external package) | 483(W)x178(D)x 550(H) |