Needle Flame Tester device meets the standard of EN/IEC 60695-11-5.
This device is designed to apply the needle flame test to the equipment, its sub-units and components, and to solid electrical insulation materials or other combustible materials.
This test device meets the standard of EN/IEC 60695-11-5.
This device is designed to apply the needle flame test to the equipment, its sub-units and components, and to solid electrical insulation materials or other combustible materials.
With the axis of the burner in the vertical position, the gas supply is adjusted so that the length of the flame is 12 mm ± 1 mm, when viewed in subdued light against a dark background.
The flame is confirmed using the apparatus and procedure detailed in Annex A. The test time for the temperature to increase from 100 °C ± 5 °C to 700 °C ± 3 °C is 23,5 s ± 1,0 s.
Needle Flame Tester Features:
– With the laser light, the distance between the sample and the burner can be adjusted.
– There are automatic or manual operating modes.
– Thanks to the automatic mode, the distance between the test specimen and the burner is automatically aligned and recorded with sensors at the end of the ignition.
– The device has a detector that detects gas leakage and ensures user safety.
– The device has a large touch screen.
– The device is supplied with the following parts:
Remote control, white pine sheet, wrapping paper, flame height meter, flue connection hose and mounted drain motor.
Description of the test apparatus
Burner | The burner to produce the test flame consist of a tube 35 mm long with a bore of 0,5 mm ± 0,1 mm and an outer diameter 0,9 mm.
The burner is supplied with butane or propane gas having a purity of at least 95 %. |
Flame | With the axis of the burner in the vertical position, the gas supply is adjusted so that the length of the flame is 12 mm ± 1 mm, when viewed in subdued light against a dark background.
The flame is confirmed using the apparatus and procedure detailed in Annex A. The test time for the temperature to increase from 100 °C ± 5 °C to 700 °C ± 3 °C is 23,5 s ± 1,0 s. |
Thermocouple | A sheathed fine wire type K with an outer sheath diameter of 0,5 mm. |
Temperature indicating, recording and timing devices | These devices is appropriate for the measurement of the time for the copper block to heat up from 100 °C ± 5 °C to 700 °C ± 3 °C with a time uncertainty of 0,1 s. The timing device has a tolerance of 0,5 s. |
Some technical parameters:
Power Supply | 220V /50Hz & 110V /60Hz |
Temperature range | 0 ~ 1000 °C |
Flame application angle | 45° or 90° |
Distance between sample and pine sheet | 200 ± 5mm |
Wrapping paper | 12 ~ 30 g / m³ |
Thickness of pine sheet | 10 mm |
Timer | 9999 x 0.1 s |
Flame application distance | 8 ± 1mm |
Flame control | lame height adjusted by valve |
Lighting | Cabin interior led lighting ≤ 20 Lx, |
Volume of chamber | 0.5 m3 |
Dimensions and weight | W: 1100 mm × D: 550 mm × H: 2000 mm; 140 kg |
Power | 1 kVA |