Features;
- Dry type or SF6 type test transformers are optional in the HV unit.
- 320 x 240 LCD screen, high speed thermal printer is optional.
- Customer can choose between AC or AC/DC testing.
- Automatic or manual mode can be selected, simple and easy operation.
- HV voltage/current, DC voltage/current, timing and results are displayed directly.
- Over voltage, over current protection. Output voltage, HV current upper limit, DC current upper limit and timing can be adjusted.
- In automatic tests, voltage increase or decrease can be stopped at any time; the test sample can be examined in more detail.
- Zero start protection. The test can be resumed when the voltage is zero again.
- Automatic timing is performed when the set voltage value is reached. After timing, the voltage is automatically reset again.
- When the HV/VG current exceeds the set value, the voltage output is automatically cut off and an alarm is given.
- During the test, the Exit button can be pressed to cancel the test. If an abnormal situation occurs, the Emergency button is pressed.
- Thanks to its interference-proof design, it can be operated under harsh electromagnetic conditions.
TECHNICAL PARAMETERS
Input Voltage | AC220V ± 10% |
Frequency | 50Hz ± 1 |
Output Voltage | 0-100kV, AC RMS |
Operation | 100% -1min, 50%-constant |
Capacity | 10kVA |
High Voltage Suplly | Oil Type Test Transformer |
Output Kontrol | Manual |
Voltage Accuracy | < 1,5% (F.S) |
Circuit Accuracy | < 1,5% (F.S) |
Time Range | 0-99H |
Measurement | Digital Screen, kV meter; AC 0-100kV, mA meter: AC 0-100mA |
Working Environment | -20℃ – 50℃; 85% RH |
Weight | Control Unit: 40 kg Oil Test Transformer: 100 kg |
Reviews
There are no reviews yet.